1. Power Supply Faults
- Low input mains voltage, poor socket contact, overload tripping of power strip.
- Power loss and aging of vehicle-mounted or built-in battery; the battery cannot drive high voltage and causes immediate shutdown.
- Emergency stop button locked after being pressed, leading to complete power-off and machine locking.
- Loose, broken or virtual connection of power cords and aviation plugs.
2. Safety Interlock Protection Faults
- Triggering of rack limit switches and pitching/lifting travel switches, resulting in automatic power-off protection.
- For some models, the cabinet door interlock is equipped; exposure is prohibited and the machine shuts down if the door is not closed tightly.
- Shutdown triggered by tilting protection and anti-collision induction of the machine.
3. High Voltage & X-ray Tube Overheating Faults
- Excessive continuous shooting times cause X-ray tube overheating protection and forced shutdown for cooling.
- Slight leakage of high-voltage modules and high-voltage cables leads to tripping of the machine’s leakage protection.
- Aging of X-ray tube filament or failure of high-voltage inverter board causes no high-voltage output and failure to work normally.
4. Control & Operation Component Faults
- Short circuit or disconnection of foot switch causes abnormal machine detection and system locking.
- Malfunction of operation panel keys and main board crash lead to program stuck and stop working.
- Fuse blown of internal circuit and power supply board.
5. Overall System Faults
- Industrial control tablet or host system crash and software error, exiting the working mode automatically.
- Internal cooling fan stop running causes high internal temperature and triggers overload protection.